A visual electrophysiology system records the electrical responses of the retina and visual pathway to light and pattern stimuli. ERG (electroretinogram) tests retinal function, VEP (visual evoked potential) tests the pathway from the eye to the visual cortex, and EOG (electro-oculogram) tests the retinal pigment epithelium. When you buy equipment, the practical choice is between a mobile system that goes to the bedside and tests infants, an all-in-one system for a dedicated test room, and an animal system for preclinical research. Check which ISCEV standard tests the system runs, how it is calibrated, and what electrodes, channels and reports come with it.
What do ERG, VEP and EOG measure?
The International Society for Clinical Electrophysiology of Vision (ISCEV) publishes the standards that most laboratories follow. Its standards cover full-field ERG, multifocal ERG, pattern ERG, VEP and EOG, each answering a different clinical question.
Swipe sideways to see all columns →
| Test | What it assesses | Typical question it helps answer |
|---|---|---|
| Full-field ERG | Generalised retinal function under dark-adapted and light-adapted conditions | Is the dysfunction in the outer or inner retina, and in the rod or the cone system? |
| Pattern ERG | Macular and ganglion cell function | Is central retinal function affected? |
| Multifocal ERG | Retinal function at many locations, so it can tell where the retina is affected | Is the loss local or widespread? |
| Flash and pattern VEP | Function of the visual pathway from eye to cortex | Is the optic nerve or the pathway behind the eye affected? |
| EOG | Retinal pigment epithelium function | Is the outer retina and pigment epithelium working? |
The ISCEV standards for full-field ERG (2022 update), pattern ERG (2024), VEP (2025), multifocal ERG (2021) and EOG (2017) are listed on the ISCEV website, together with guidelines for calibration of stimuli and recording instruments (2023) and extended protocols such as the photopic negative response and S-cone ERG. Which tests a clinic needs depends on its patients: a children’s hospital, a neuro-ophthalmology unit and a retina clinic will not order the same configuration.
Mobile, all-in-one or animal system?
Visual electrophysiology equipment comes in three common layouts. Choose by where the subject will be tested.
- Mobile system: amplifier, stimulators and computer on a movable table, with a light flash stimulator that can be taken to a lying subject or an infant. It suits wards, children’s hospitals and sites that cannot move the patient to a fixed room.
- All-in-one system: amplifier, flash and pattern stimulators in a single host, so the subject stays in one position for every test. It suits a dedicated electrophysiology room with a regular workload.
- Animal system: a platform with a heated blanket, electrode holder and flash stimulator for mice, rats and other species, for preclinical studies and disease-model research. It is for research use.
What to compare when you buy
- Tests and protocols: confirm in writing which ISCEV standard protocols (full-field ERG, pattern ERG, VEP, EOG) and extended protocols are included, and whether you can write your own programs.
- Stimulators: flash colours and intensity range, background luminance, pattern types and stimulus fields. A full-field dome (Ganzfeld-type) or flash bowl is needed for full-field ERG; a pattern stimulator is needed for pattern ERG and pattern VEP.
- Amplifier: channel count (the RetiMINER-C has two as standard, with five or eight optional), A/D resolution, input noise, filters and whether it runs from a battery.
- Calibration: ISCEV expects the user to check calibration, with stimulus strength measured at the eye. Ask what calibration tools, instructions and service the supplier offers.
- Reference values: ISCEV notes that reference ranges depend on the electrode type, and that each laboratory should ideally build its own. Ask whether the supplier gives starting reference data, and plan to verify it locally.
- Electrodes and consumables: which electrode types are supplied, which are optional and how they are re-ordered.
- Subject handling: seated, lying and infant testing, infrared monitoring, dark and light adaptation timers, and chin or head support.
- Software and reports: offline averaging, manual rejection of artefacts, export formats, network sharing and DICOM, and the report languages you need.
- Upgrade path: whether multifocal ERG or other modules can be added later.
- Room, training and documents: the test-room layout and lighting, operator training, and the registration and documentation your market requires.
ENOCHTEC visual electrophysiology systems
Figures below are from the ENOCHTEC product pages; the exact configuration, electrodes and documents for your market are confirmed in the quotation.
Swipe sideways to see all columns →
| RetiMINER-C | RetiMINER-P | RetiMINER Animal | |
|---|---|---|---|
| Layout | Mobile system on a movable motorized table | All-in-one integrated host | Animal platform with cantilever flash stimulator |
| Tests | PVEP, FVEP, PERG, FERG, EOG; S-cone ERG, PhNR, double flash ERG, On/Off ERG | Pattern VEP, flash VEP, pattern ERG, flash ERG, EOG | Predefined pattern VEP, flash VEP, pattern ERG and flash ERG protocols; custom protocols |
| Subject position | Seated, lying; infants | Single position for all tests | Anaesthetised animals on a heated platform |
| Amplifier | 24-bit DC, battery powered; 2 channels, 5 or 8 optional | 24-bit, optically isolated USB; input noise ≤4 µVpp | 24-bit, optically isolated USB; input noise ≤4 µVpp |
| Flash stimulator | Under 3 kg; 100 cd·s/m² in under 4 ms; range 0–1000 cd·s/m² | 30 cd·s/m² in under 4 ms; range 0–100 cd·s/m²; visor can serve as a dome | 100 cd·s/m² in under 4 ms; range 0–1000 cd·s/m²; 90° cantilever |
| Pattern stimulator | LCD, contrast above 98% | Contrast above 98%; checkerboard, stripes and other fields | Optional |
| Upgrades and options | Multifocal ERG and VEP upgrade; 5- or 8-channel amplifier; infrared and ultraviolet; animal test set | DICOM and electrode sets as options | Ultraviolet, pattern stimulator and DICOM options |
| Best for | Bedside, infants and mobile service | Dedicated test room | Preclinical research (mice, rats; other species) |
For related equipment, see Visual Field Analyzer and Perimeter Guide and OCT vs Fundus Camera, or plan a full ophthalmology clinic.
Frequently asked questions
What is the difference between ERG and VEP?
ERG records the electrical response of the retina to light or pattern stimuli. VEP records the response of the visual cortex, so it reflects the whole visual pathway from the eye through the optic nerve to the brain.
What does an EOG test?
EOG tests the function of the retinal pigment epithelium and the outer retina by recording the standing potential of the eye during eye movements between fixation lights.
What are the ISCEV standards?
They are the International Society for Clinical Electrophysiology of Vision standards for full-field ERG, multifocal ERG, pattern ERG, VEP and EOG, plus guidelines and extended protocols, which define the minimum stimuli, recording and reporting for each test.
Can an ERG or VEP system test infants and bed-bound patients?
Mobile systems are designed for it. The ENOCHTEC RetiMINER-C has a flash stimulator under 3 kg and a movable motorized table for testing seated and lying subjects and infants at the bedside.
Do I need a mobile or an all-in-one system?
Choose a mobile system if patients cannot come to a fixed room, and an all-in-one system if you have a dedicated test room. The RetiMINER-P keeps the subject in one position for pattern, flash and EOG tests.
Can the same system be used for animal research?
The RetiMINER-C offers an optional animal test set and platform, and the separate RetiMINER Animal system is built for mice and rats, with options for other species. Both are for research use.
Do reference values come with the equipment?
Reference ranges depend on the electrode type, and ISCEV advises each laboratory to establish its own and to verify any external data before use. Ask the supplier what starting data and training are provided.
Looking for an ERG and VEP system?
Tell us whether you test at the bedside, in a dedicated room or in animals, which tests you need, and your local power standard. We will recommend a configuration and send a quotation.
Request a quoteSources: ISCEV: Standards, guidelines and extended protocols; Robson et al., ISCEV Standard for full-field clinical electroretinography (2022 update), Documenta Ophthalmologica. Device figures are from the ENOCHTEC product pages.
This guide is general information for equipment buyers. Performing and interpreting visual electrophysiology tests and diagnosis are the responsibility of qualified eye-care professionals.